[KCH606] life-support machine (portable) is creates the usable life-support machine by the Zibo Kechuang
medical instrument limited company develops, this machine is air
operated, the microcomputer control life-support machine, this
machine mainly has function and so on IPPV, SIMV, CPAP, SIGH,
manually controlled breath, may through the hypothesis breath
frequency, attract shouts compared to and adjusts the moist spirit
realization to patient's breath management, this machine major
component uses the high grade import product, enhanced this machine
reliability and the stability greatly, this machine reserve lithium
battery may work for 10 hours, if the battery exhausts may the
manually controlled breath. This machine structure reasonable, the
ease of operation, the performance are reliable, the exchange, the
battery, radio may use.
Main performance and technical specification
- Ventilates the pattern: IPPV, SIMVf/2, SIMVf/4, CPAP1, CPAP2, SIGH,
manually controlled breath.
- Air course performance:
A) air course most Da'an full pressure:Is not bigger than 6Kpa.
B) inspires the oxygen concentration: Is smaller than 45%.
3. Ventilates the performance:
A) breathes the frequency: 6 ~ 60/minutes, error ±15%.
B) attracts shouts the ratio: 1:1, 1:2, 1:3, 1:4, 1:5, 2:1, 2:3,
3:5, error ±15%.
C) triggers the pressure:The inspiration triggers the pressure in
-2KPa ~ -1Kpa, the allowable deviation ±20%; The inspiration
triggers the pressure in -1KPa ~ -0.4Kpa, allowable deviation
±200Pa; The inspiration triggers the pressure in -0.4KPa ~
-0.1C$kpa, allowable deviation ±50Pa.
D) moist spirit: 300 ~ 1000ml; Error ±20%.
E) the minute passes the spirit: Is not smaller than 18 L/min. - Reports to the police the performance:
A) the pressure upper limit reports to the police: 2 ~ 6KPa, the
error ±20%, the acousto-optics reports to the police.
B) the low pressure limit reports to the police: 0 ~ 2KPa, the
acousto-optics reports to the police.
C) the battery owes the pressure: The acousto-optics reports to the
police.
D) static sound:Is not bigger than 2 minutes. - Monitor performance:
A) moist spirit: 0 ~ 1200ml. 0 ~ 300ml, error ±60ml; 300 ~ 1200ml
error ±20%.
B) actual frequency.
C) gas channel pressure: Liquid crystal display.
D) exchanges, the charge, reports to the police, triggers the
instruction. - Entire machine performance:
A) power source:Exchange 220V±22V.50Hz±1Hz.
B) entire machine noise: Is not bigger than 65dB (A).
C) system compliance: Is not bigger than 4ml/100Pa.
D) lithium battery: 7.2V 1800mAh.
E) radio may the manually controlled breath.
F) weight 2.5kg.